YAMARI Việt Nam

Our Calibration equipment

Thứ Sáu, 12/12/2025
Classification Working standard/calibration equipment Calibration temperature(℃) Product under calibration
Fixed point calibration Triple point apparatus of mercury -38.8344 Platinum resistance thermometer, thermocouple, and thermometer with indicator
Freezing point equipment 0
Triple point device for water 0.01
Gallium melting point apparatus 29.7646
Indium freezing point equipment 156.5985
Tin freezing point device 231.928
Zinc freezing point equipment 419.527
Aluminum freezing point equipment 660.323
Silver freezing point equipment 961.78
Copper freezing point equipment 1084.62 Thermocouple and thermometer with indicator
Comparative calibration Platinum resistance thermometer/liquid bath or dry block bath -196~250 Platinum resistance thermometer, thermocouple, and thermometer with indicator
Platinum resistance thermometer/powder bath 250~660
R thermocouple/tubular furnace or spherical furnace 200~1554 Thermocouple and thermometer with indicator
Platinum resistance meter or R thermocouple -100~1100 Thermometer calibration device

※Glass thermometers should be calibrated using freezing equipment and liquid or powder baths.

◆Special calibration equipment◆

Cryokelvin
Top load type refrigerator cryostat. Used to achieve any temperatures between-196°C and-80°C in JCSS calibration. The lower limit temperature can be achieved up to 20K(-253℃), and is also evaluated at liquid-hydrogen temperatures.

Liquid-helium tank 250L
The sensor is immersed in liquid-helium to evaluate 4K(-269℃).

Ultra-high-temperature thermocouple calibration furnace
Realizes 1100℃ to 2000℃. Type C we sell is used for lot-testing of tungsten-rhenium thermocouples (reference thermocouples are NIST traced).

Viết bình luận của bạn
Hotline Chat zalo E-mail